The blood gas
analyzer is used for the clinical detection of the three programs. measuring the
blood is commonly used in medical
devices, these three indicators that: ph, p02, pCO2. Brief
introduction of the following Plantier medical device manufacturers experts
blood gas analyzer liquid road system, as follows:
Blood gas analysis
instrument of Fluid Systems installed a variety of liquid containers, connecting
pipes, solenoid valve and vacuum pumps and peristaltic pump composed.
Buffer calibration
process: buffer 1 or 2 in the peristaltic pump suction inlet location after
conversion plate to reach the measurement chamber; rinse in the negative
pressure suction of waste bottle , the liquid flows through the washing fluid -
converter position - inlet - Measure Room - peristaltic pumps - waste bottle.
Thus, the complete flushing of the measuring chamber and the entire
pipeline.
2:00 each
calibration, the blood gas analyzer with cleaning fluid automatically cleaned
once a measurement chamber. This process is peristaltic pump suction, the
process is: the cleaning solution - the location of the conversion plate - Inlet
- the measurement chamber - peristaltic pumps - waste bottle. Measurement, the
conversion plate is stopped at the gap. Sample peristaltic pump suction inlet
input into the measurement chamber.
All above are
Introduction of blood
gas analyzer liquid road system.