Full wavelength elisa reader is extremely used throughout the country, because
of the instrument performance, stability, and accuracy of data results, more
importantly. The instrument has a robust detection function. Its range of
applications across a variety of industries. The most commonly used in the food
industry and the medical
industry.
Full wavelength
microplate reader offer price how much?
The Perlong
Medical is a professional production of full wavelength microplate reader
manufacturers and manufacturers. The Prang medical production of
full wavelength microplate reader instrument measurement accuracy, good
repeatability, quality control function, no external computer can automatically
provide 31 days QC data and X, SD, CV values; the instrument preset operating
procedures, test results can be stored; boot after the self-test function, to
ensure the instrument each time work in normal working condition; and a wide
range of instruments to measure the performance indicators have reached the
international advanced level. Color LCD with touch-screen operation, can do
genetic testing, display calibration curve advantages.
The following brief
full wavelength microplate reader often malfunction and processing
methods.
Phenomenon of a
fault
The washer head clogging:
Failure Analysis and Maintenance: sample
fibrin. Instrument self-test or washer process. Instrument self-test, can shut
down the washer head treatment, re-POST: If the fault is in the washer process,
you need to continue to press the "pause" key of the instrument, processing
washer head wash board.
Processing method: Remove the washer head. Careful
observation of each of the liquid injection port, and a liquid suction port. If
you found a fibrin clot. Picked to go with a disposable syringe needle fibrin
clot, and then clean the washer head; found no fibrin clot. Can be directly on
the the Microplate head cleaning. When cleaning the washer head 5ml disposable
syringe without needle, draw the the total liquid suction port of the distilled
water alignment Xiban head or total fluid intake. Blocked another exit, the
liquid inside the disposable syringes forcibly discharged. Total fluid outlet or
the total fluid intake alternately repeated washing, to complete the cleaning of
the Microplate head clogging. When the temperature is low. 5O ℃ 6o ℃ distilled
water heated to rinse the effect is more obvious.
Symptom 2
Sample syringe
and silicone tube junction leakage or off:
Failure Analysis and Maintenance:
Because the sample needle clogging or liquid waste pipe is blocked, resulting in
increased pressure of the pipeline, the silicone tube to break or fall off.
Instrument self-test or the test process.
Processing methods:
to stop instrument self-test or test, remove needle, very fine copper wires
inside Tonga syringe. Check the silicone tube if there is rupture phenomenon,
replace broken, connection. After the passage of the power-on self-test
instrument pipeline flush. Troubleshooting, start the test, twice a week to
remove the sample needle soaked in 5% ~ 1O% sodium hypochlorite solution for 2h
and pipe with a 5% sodium hypochlorite solution rinse instruments. You can
reduce the occurrence of this failure.
The failure phenomenon POST Times
reagent disk error. The reagent disc kept turning, unable to stop.
Fault
Analysis and Maintenance: This is caused due to the reagent disc underside of a
small magnet off. The analyzer reagent disk using magnetic sensors to locate, at
the bottom surface of the reagent disk set with the L small pieces magnet.
Sensor fitted in the instrument underlying the above position reagent tray lined
with small magnets to the above the sensor, the sensor will send a signal to
locate the reagent disc, so the instrument self-test will pass.
Processing
methods: shutdown. Picked reagent disk, reinstall the small magnet back, clean
the sensor. The power-on self-test, the machine is operating
normally.
Through the
introduction of the above content, I believe you have some understanding of the
full wavelength elisa
reader, especially the instrument failed after processing. The
Pulang medical production of medical device products, superior quality.
Including: wavelength microplate reader, hematology analyzer, urine analyzer,
biochemistry analyzer, coagulation analyzer, washer, hemoglobin analyzer and
other equipment.
If you have any
questions or you want to consult the full wavelength microplate reader offer
price, you can call the the the Prang medical free hotline :400 -6656-888
contact our customer service staff. We are happy to serve you.