FIA8000 Immunoassay Analyzer

Application: FIA8000 Immunoassay Analyzer is designed to support rapid
test kit for cardiac markers, it tells the accurate value of the analytes
and is a very important contributor to the diagnosis and treatment of
cardiovascular diseases. It is warmly welcomed in clinics, emergency
room, operation room, ICU,CCU etc.

Test Items: cTnI, NT-proBNP, NT-proBNP/cTnI, CK-MB/cTnI/Myo, hs-CRP
D-Dimer

Sample: Serum/Plasma/Whole Blood



Background Count: The voltage must be higher than 3500mV when test a blank
quanlity control card.
Linearity: r≥0.99(0--4000mV)
Repeatability: CV≤1%
Stability: The voltage variation should be less than ±2% if test the same
quanlity control card within in 1h.
Working Environment: Temperature: (+15°C~+35°C)



Humidity: 10%~85%
Atmospheric Pressure:70.0kPa~106.0kPa
Power Supply:AC100V~240V 50Hz~60Hz
Transport and Storage Environment: Temperature: (-10°C~+40°C)
Humidity: ≤93%
Atmospheric Pressure:50.0kPa~106.0kPa