Introduction of Blood Gas Analyzer Liquid Road System

The blood gas analyzer is used for the clinical detection of the three programs. measuring the blood is commonly used in medical devices, these three indicators that: ph, p02, pCO2. Brief introduction of the following Plantier medical device manufacturers experts blood gas analyzer liquid road system, as follows:
Blood gas analysis instrument of Fluid Systems installed a variety of liquid containers, connecting pipes, solenoid valve and vacuum pumps and peristaltic pump composed.
Buffer calibration process: buffer 1 or 2 in the peristaltic pump suction inlet location after conversion plate to reach the measurement chamber; rinse in the negative pressure suction of waste bottle , the liquid flows through the washing fluid - converter position - inlet - Measure Room - peristaltic pumps - waste bottle. Thus, the complete flushing of the measuring chamber and the entire pipeline.
2:00 each calibration, the blood gas analyzer with cleaning fluid automatically cleaned once a measurement chamber. This process is peristaltic pump suction, the process is: the cleaning solution - the location of the conversion plate - Inlet - the measurement chamber - peristaltic pumps - waste bottle. Measurement, the conversion plate is stopped at the gap. Sample peristaltic pump suction inlet input into the measurement chamber.
All above are Introduction of blood gas analyzer liquid road system.