Full wavelength microplate reader offer price | full wavelength microplate reader manufacturers

Full wavelength elisa reader is extremely used throughout the country, because of the instrument performance, stability, and accuracy of data results, more importantly. The instrument has a robust detection function. Its range of applications across a variety of industries. The most commonly used in the food industry and the medical industry.
     Full wavelength microplate reader offer price how much?
     The Perlong Medical is a professional production of full wavelength microplate reader manufacturers and manufacturers. The Prang medical production of full wavelength microplate reader instrument measurement accuracy, good repeatability, quality control function, no external computer can automatically provide 31 days QC data and X, SD, CV values; the instrument preset operating procedures, test results can be stored; boot after the self-test function, to ensure the instrument each time work in normal working condition; and a wide range of instruments to measure the performance indicators have reached the international advanced level. Color LCD with touch-screen operation, can do genetic testing, display calibration curve advantages.
The following brief full wavelength microplate reader often malfunction and processing methods.
Phenomenon of a fault
The washer head clogging:
Failure Analysis and Maintenance: sample fibrin. Instrument self-test or washer process. Instrument self-test, can shut down the washer head treatment, re-POST: If the fault is in the washer process, you need to continue to press the "pause" key of the instrument, processing washer head wash board.
Processing method: Remove the washer head. Careful observation of each of the liquid injection port, and a liquid suction port. If you found a fibrin clot. Picked to go with a disposable syringe needle fibrin clot, and then clean the washer head; found no fibrin clot. Can be directly on the the Microplate head cleaning. When cleaning the washer head 5ml disposable syringe without needle, draw the the total liquid suction port of the distilled water alignment Xiban head or total fluid intake. Blocked another exit, the liquid inside the disposable syringes forcibly discharged. Total fluid outlet or the total fluid intake alternately repeated washing, to complete the cleaning of the Microplate head clogging. When the temperature is low. 5O ℃ 6o ℃ distilled water heated to rinse the effect is more obvious.
Symptom 2
Sample syringe and silicone tube junction leakage or off:
Failure Analysis and Maintenance: Because the sample needle clogging or liquid waste pipe is blocked, resulting in increased pressure of the pipeline, the silicone tube to break or fall off. Instrument self-test or the test process.

Processing methods: to stop instrument self-test or test, remove needle, very fine copper wires inside Tonga syringe. Check the silicone tube if there is rupture phenomenon, replace broken, connection. After the passage of the power-on self-test instrument pipeline flush. Troubleshooting, start the test, twice a week to remove the sample needle soaked in 5% ~ 1O% sodium hypochlorite solution for 2h and pipe with a 5% sodium hypochlorite solution rinse instruments. You can reduce the occurrence of this failure.
The failure phenomenon POST Times reagent disk error. The reagent disc kept turning, unable to stop.
Fault Analysis and Maintenance: This is caused due to the reagent disc underside of a small magnet off. The analyzer reagent disk using magnetic sensors to locate, at the bottom surface of the reagent disk set with the L small pieces magnet. Sensor fitted in the instrument underlying the above position reagent tray lined with small magnets to the above the sensor, the sensor will send a signal to locate the reagent disc, so the instrument self-test will pass.
Processing methods: shutdown. Picked reagent disk, reinstall the small magnet back, clean the sensor. The power-on self-test, the machine is operating normally.

Through the introduction of the above content, I believe you have some understanding of the full wavelength elisa reader, especially the instrument failed after processing. The Pulang medical production of medical device products, superior quality. Including: wavelength microplate reader, hematology analyzer, urine analyzer, biochemistry analyzer, coagulation analyzer, washer, hemoglobin analyzer and other equipment.
If you have any questions or you want to consult the full wavelength microplate reader offer price, you can call the the the Prang medical free hotline :400 -6656-888 contact our customer service staff. We are happy to serve you.

Article : From Perlong medical (http://www.pl999.net/)