Problems occur in use about hematology analyzer

Hematology analyzer in the operating process often meet such situations, such as: white blood cell problems, problems hemoglobin, white blood cells, red blood cells, platelets issues and quality control problems. The Perlong the medical instruments wangbian specific order as follows:
· Hematology analyzer operation leukocytes.
Reagent temperature, if the temperature is too high, the white blood cells to be low hemolysin hemolytic excessive, when the temperature dropped to 25 degrees, if the white blood cells is too high, the general temperature is too low, then the temperature is raised to 25 degrees, there is a situation, that is, white blood cells is too high, then you may the hemolysin pipeline bubbles inside, hemolytic inadequate services, system assistant menu inside, the replacement of the hemolysin can be. There is blood analyzer plugging holes: plugging holes when leukocytes deviation is too large, you should take the maintenance.
· Hematology analyzer operating in the hemoglobin problem.
If hemoglobin is low or high, this time in the service - waveform display - inside the key, look on the 8th HGB voltage If the voltage is not 4.2V, generally the voltage adjusted to 4.2V (debug: turn the machine on the left side panel , see the bottom of the knob, turn the knob to clockwise to increase the voltage, and vice versa, reducing the voltage, debugging, monitoring voltage waveform display inside, until the a 4.2V date), if in the 4.1 - 4.3 between can also be adjusted voltage parameters according to the target value of the quality control of blood HGB debugging HGB.
· Blood analyzer operating in the white blood cells, red blood cells, platelets issues.
If the white blood cells, red blood cells, platelets deviation is large, the first enzyme soak count pool.
· Hematology analyzer operating in the quality control problems.
Maintenance after the end of the process, if the white blood cells, red blood cells, platelets also deviation, this time to do quality control: do quality control ready condition, first bottle quality control of blood, and the target value of the quality control of blood, with the quality control of blood in our machine above measurement, if the measured results and the target value is very close, then the machine work is normal, if the result of the measurement and a target value comparison, the difference is great, indicating that the machine has a drift, this time to debug the machine parameters (debug, first menu, then settings, system settings, password management, and then enter the password 2008 and press OK to return to the interface, directly press the 4 key inside parameter debug), tune the parameters and the proportion of cells positive relationship that the measured results low quality control target values, simply increase the parameters. On the contrary, no exception. Results and the target value until the call up close so far.
Blood cell analyzer for your reference learning more about coagulation analyzer products information, please refer to the official website of the Perlong medical devices.

Article : From Perlong medical (http://www.pl999.net/)